National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Breakthrough (effluent = 5% of test concentration) occurred after sampling for 60 min at 0.046 L/min from an atmosphere containing 14500 mg/m3 1,1,2-trichloro-1,2,2-trifluoroethane in dry air. Desorption efficiency for 18 samples of SKC lot 105 activated coconut charcoal spiked with 5.6 to 23 mg 1,1,2-trichloro-1,2,2-trifluoroethane averaged 1.01 with = 0.02.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Carbon disulfide is toxic and a serious fire and explosion hazard (flash point = –30 °C) . 1,1,2-Trichloro-1,2,2-trifluoroethane is a narcotic [4] . Work with these compounds only in a hood. SAMPLING: 1.
Calibrate each personal sampling pump with a representative sampler in line. 2.
Break the ends of the sampler immediately before sampling. Attach sampler to personal sampling pump with flexible tubing. 3.
Sample at an accurately known flow rate between 0.01 and 0.05 L/min for a total sample size of 0.1 to 3 L. 4.