“ Breakthrough cap acities were de term ined in dry air. Storage stability was assessed at 7, 14, and 30 days. Measurement precisions (Table 4) were determined by spiking sampling media with amounts corresponding to one-half, one, and two times the OSH A standard for nominal air volumes. Desorption efficiencies for spiked sam plers con taining only one co m pou nd e xce ede d 75 % [2,3,4,8 ] . Issue 3: The desorption efficiency, at levels ranging from 10 times the LOQ to 0.1 times the REL, was determined by spiking known am oun ts of analytes (in CS 2) on coconut shell charcoal tubes. ”