National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Transfer the liquid from the bubbler, quantitatively, to a graduated cylinder. Bring volume to 15 mL with distilled water. CALIBRATION AND QUALITY CONTROL: 9. 10. Prepare working standards (20 to 800 μg/10 mL) by adding appropriate aliquots of calibration stock solution to distilled water. Analyze working standards together with samples and blanks (steps 11 through 13) . Prepare a calibration graph of area vs. μg of maleic anhydride per 15 mL of sample. ”