National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Atomize cycle: 10 sec at 2700 to 3000 °C; atomization is too slow for reproducible results below 2600 °C; "maximum power" feature will increase sensitivity. Inject sample aliquot manually or with autosampler. Measure peak height. Analyze a series of standards before and after each set of samples and analyze one mid-range standard after each ten samples.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Mean recovery was 93% (24 samples) in the range 5 to 121 μg/m 3 (0.0007 to 0.017 ppm) . Breakthrough (effluent = 5% of test concentration) did not occur after sampling for 240 min at 0.475 L/min from an atmosphere containing 34 μg/m 3 (0.005 ppm) Ni (CO) 4 in air at 22 °C and 19% RH. Desorption efficiency for eighteen samples containing 0.08 to 0.5 μg Ni (nickel nitrate standard solution added to 120 mg charcoal) averaged 0.934 with S r = 0.029. Nickel carbonyl was observed to pass quantitatively through 0.8-μm cellulose ester membrane filters in these experiments.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

The prefilter must be used when particulate nickel is present in the air to be sampled. Connect prefilter to sorbent tube with a short section of plastic tubing. 2. Personal sampling pump, 0.05 to 0.2 L/min, with flexible connecting tubing. 3. File, triangular. 4. Atomic absorption spectrophotometer equipped with: a. Heated graphite atomizing rod, tube or furnace (either non-pyrolitic or pyrolitic; better sensitivity with pyrolitic) . NOTE: Reproducible control of times and temperatures during dry, char, and atomize cycles is essential
Source: Wikisource