National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Atomize cycle: 10 sec at 2700 to 3000 °C; atomization is too slow for reproducible results below 2600 °C; "maximum power" feature will increase sensitivity. Inject sample aliquot manually or with autosampler. Measure peak height. Analyze a series of standards before and after each set of samples and analyze one mid-range standard after each ten samples. ”