National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Prepare a calibration graph of peak height vs. amount (μg) of fluoride per 20 mL of sample. (Optional) . Determine recovery (R) for each lot of tubes used for sampling in the concentration range of interest. Prepare four tubes at each of five levels plus three media blanks. a. Collect a known amount of SO 2F2 gas onto each charcoal tube (steps 1 through 9) . b. Analyze samples in the same manner as field samples (steps 12 through 14) . c. Prepare graph of recovery vs. μg sulfuryl fluoride. MEASUREMENT: 12. 13. 14. Set ion chromatograph to conditions given on page 6012-1. ”
