National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Wear protective equipment (gloves, labcoats, and safety glasses) when working with HF. Do not use HF with glass labware. Perform all acid digestions in a fume hood. SAMPLING: 1. 2.
Calibrate each personal sampling pump with a representative sampler in line. Sample at an accurately known flow rate between 1 and 4 L/min for a total sample size of 200 to 1000 L. Do not exceed 2 mg total dust loading on the filter.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

ESTIMATED LOD: 50 μg soluble W per sample; 125 μg insoluble W per sample [2] PRECISION (Sr) :
0.029 [1]
APPLICABILITY: The working range of this method is 0.25 to 5 mg/m 3 (soluble) and 0.6 to 5 mg/m 3 (insoluble) for a 400-L air sample. This is an elemental analysis and not compound-specific. For cemented tungsten carbide samples, cobalt can be quantitatively determined from the HCl extract used to remove interferences.
INTERFERENCES: None known. Ni, Mo, V, Mn, Cr, Co and Fe do not interfere at concentrations up to 50 times the tungsten concentration [3] .
Source: Wikisource