National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Add exactly 15 mL Girard T solution to each bubbler using a 15-mL pipet. Mark the initial liquid level in the bubbler with a glass marker. Make impinger-to-trap and trap-to-sampling pump connections with flexible inert tubing. 3.
Sample at an accurately known flow rate between 0.1 and 0.5 L/min for a total sample size of 6 to 60 L. NOTE:
Higher flow rates will cause frothing of the collection medium. If amount of liquid condensed in the trap is greater than 1 mL, collection efficiency of bubbler may be reduced and sample may be invalid.
”
Source: Wikisource

National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Tap bubbler stem lightly against bubbler body to drain contents into the body. If necessary, bring samples up to the 15-mL mark with distilled water. Swirl bubbler to mix contents well. Do not add solution collected in the trap to the sample. 5.
Transfer a 5-mL aliquot to a 100-mL flask and bring to volume with HPLC mobile phase. CALIBRATION AND QUALITY CONTROL: 6.
Calibrate daily with at least six working standards over the range 0.007 to 4 mg acetaldehyde per mL (0.1 to 60 mg acetaldehyde per sample) . a.
”
Source: Wikisource