National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ The concentration of acetic acid was independently verified by a total hydrogen analyzer. Desorption efficiency was 0.96 in the range 2.1 to 8.4 mg per sample. Breakthrough (5% on back section) was never achieved and testing was discontinued after 4.6 hrs when 10.4 mg of acetic acid was collected without breakthrough for a 269-L sample at 90% RH. A user check gave an estimated LOD of 0.01 mg per sample and a desorption efficiency of 1.01 in the range 0.3 to 5 mg per sample [2] . ”