National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

The concentration of acetic acid was independently verified by a total hydrogen analyzer. Desorption efficiency was 0.96 in the range 2.1 to 8.4 mg per sample. Breakthrough (5% on back section) was never achieved and testing was discontinued after 4.6 hrs when 10.4 mg of acetic acid was collected without breakthrough for a 269-L sample at 90% RH. A user check gave an estimated LOD of 0.01 mg per sample and a desorption efficiency of 1.01 in the range 0.3 to 5 mg per sample [2] .
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Care should be taken to avoid skin contact with formic acid and/or acetic acid. These reagents may cause severe burns.
SAMPLING: 1. 2. 3. 4.
Calibrate each personal sampling pump with a representative sampler in line. Break the ends of the sampler immediately before sampling. Attach sampler to personal sampling pump with flexible tubing. Sample at an accurately known flow rate between 0.01 and 1 L/min for a total sample size of 20 to 300 L. Cap the samplers and pack securely for shipment.
SAMPLE PREPARATION: 5. 6. 7.
Source: Wikisource