National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Add calibration stock solution with a microliter syringe to ethyl ether in a 10-mL volumetric flask and dilute to the mark. Analyze together with samples, blanks and control samples (steps 10 through 12) . Prepare calibration graph of peak area vs. mass (mg) of benzoyl peroxide per sample. Determine recovery of benzoyl peroxide from filters at least once for each lot of samplers in the calibration range (step 7) . Prepare three filters at each of five levels plus media blanks. Add aliquot of calibration stock solution with a microliter syringe directly onto a media blank filter in a vial. ”