National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Chlordane is a potential human carcinogen [5] . SAMPLING: 1. Calibrate each personal sampling pump with a representative sampler in line. 2. Assemble the sampler and connect the sorbent tube (ends broken just before the connection) to the sampling pump with backup section nearest the sampling pump. Keep the sampler in a vertical position during sampling. 3. Sample at an accurately known flow rate between 0.5 and 1 L/min for a total sample size of 10 to 200 L. 4. Remove the sorbent tube from the outlet of the cassette and connect it to the inlet side of the cassette. ”