National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Prepare calibration graph (sum of areas of selected peaks vs. μg chlorinated diphenyl oxide) . Determine recovery (R) at least once for each batch of filters used for sampling in the calibration range. Prepare three filters at each of five levels plus three media blanks. a. Deposit a known amount of calibration stock solution onto the filter. Allow filters to air dry. b. Store samples overnight in vials. c. Prepare (step 4) and analyze with working standards (steps 8 and 9) . d. Graph R vs. μg chlorinated diphenyl oxide.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Prolonged skin contact with chlorinated diphenyl oxide can cause chloracne; acute and chronic exposure can cause liver damage [3,4] .
SAMPLING: 1. 2. 3.
Calibrate each personal sampling pump with a representative sampler in line. Sample at an accurately known flow rate between 0.5 and 1.5 L/min for a total sample size of 8 to 200 L. Within 1 h of sampling, using tweezers, carefully transfer the filter and backup pad to a vial.
Source: Wikisource