National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Determine the mass, μg (corrected for R) of Ribavirin found in the sample (W) , and in the average media blank (B) , from the calibration graph. Calculate concentration, C, of Ribavirin in the air volume sampled, V (L) : EVALUATION OF METHOD: This method is a modification of a bulk assay procedure developed by Eastman Kodak Company [3] . Measurement precision, Sr, was 0.057 with average recovery of 100% representing no bias, based on 16 samples ranging from 19.2 to 112 μg per filter. Sampling precision was not determined. ”