Cresol and phenol cause severe burns [5] . They are toxic if absorbed through skin, inhaled or ingested. All work with them should be performed in a hood.
SAMPLING: 1. 2. 3. 4.
Calibrate each personal sampling pump with a representative sampler in line. Break the ends of the sampler immediately before sampling. Attach sampler to personal sampling pump with flexible tubing.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Place the front sorbent section (together with the front glass wool plug) and back sorbent sections of the sampler tube in separate vials. Discard the other plugs. Add 2.0 mL methanol to each vial. Attach crimp cap to each vial. Ultrasonicate 30 min.
CALIBRATION AND QUALITY CONTROL: 8.
9.
Calibrate daily with at least six working standards over the range 1 to 800 μg cresol and phenol per sample. a. Add known amounts of calibration stock solution to methanol in 10-mL volumetric flasks and dilute to the mark. b. Analyze together with samples and blanks (steps 11 and 12) . c.
Source: Wikisource