National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Carbon disulfide is toxic and extremely flammable (flash point = -30 °C) ; work with it only in a hood. Ethylene chlorohydrin is highly toxic, has poor warning properties, and as a liquid, is readily absorbed through the skin and penetrates rubber [3] . Toxic effects may be cumulative. Wash liquid from skin immediately if contacted. SAMPLING: 1. 2. 3. 4.
Calibrate each personal sampling pump with a representative sampler in line.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Prepare calibration graph (ratio of peak area of analyte to peak area of internal standard vs. μg ethylene chlorohydrin) . Determine desorption efficiency (DE) at least once for each lot of charcoal used for sampling in the calibration range (step 8) . Prepare three tubes at each of five levels plus three media blanks. a. Remove and discard the back sorbent section of a media blank sampler. b. Inject a known amount (2 to 20 μL) of calibration stock solution, or a dilution thereof, directly onto front sorbent section with a microliter syringe. c. Cap the tube. Allow to stand overnight. d.
Source: Wikisource