Prepare calibration graph (peak area vs. mg nitromethane) . Determine desorption efficiency (DE) at least once for each lot of sorbent used for sampling in the range of interest. Prepare three tubes at each of five levels plus three media blanks. a. Remove and discard back sorbent section of a media blank sampler. b. Inject a known amount (2 to 20 μL) of calibration stock solution directly onto front sorbent section with a microliter syringe. c. Cap the tube. Allow to stand overnight.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Nitromethane vapor forms highly explosive mixtures with air. When highly confined and heated, it can detonate [3] .
SAMPLING: 1. 2. 3.
4.
Calibrate each personal sampling pump with a representative sampler in line. Break the ends of the sampler immediately before sampling. Attach sampler to personal sampling pump with flexible tubing. Sample at an accurately known flow rate between 0.01 and 0.05 L/min for a total sample size of 1.2 to 3 L. NOTE: High pressure drop across the sampler may occur at flow rates above 0.05 L/min, giving inaccurate sample volumes.