National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

C. I. Direct Brown 95, 24.3 to 243 μg dye per sample, 78.5% recovery ( Sr = 6.7%) ; C. I. Direct Black 38, 12.5 to 250.0 μg dye per sample, 78.1% recovery ( Sr = 12.8%) . The pooled precisions ( Sr) for samples at three concentrations in the measurement range for each dye are as follows: Direct Red 28, 0.045; Direct Blue 6, 0.061; Direct Brown 95, 0.072; Direct Black 38, 0.078. Since the method does not distinguish between various benzidine-based dyes, recovery correction factors cannot be applied unless a single known dye is contained on the filter.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Prepare three filters at each of five concentration levels. c. Analyze the filters by steps 3 through 9 and 12 through 14. d. On the day of analysis of the filters, add the same volume of dye solution which was spiked onto the filters to 2.0 mL H 2O. Reduce this solution (steps 7 through 9) and analyze together with samples and standards (steps 12 through 14) . e. Calculate recovery (μg benzidine, o-tolidine, or o-dianisidine recovered from the filter/μg from step 11.d) . f. Prepare graph of recovery vs. μg benzidine, o-tolidine or o-dianisidine recovered.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

The metho d will not distinguish benzidine obtained from reduction of the dyes and free benzidine in the dye formulation.
INTERFERENCES: Aniline, azobenzene, p-aminophenol, p-phenylenediamine or p-nitroaniline do not interfere in the measurement when present in equimolar amounts.
OTHER METHODS: This revises P&CAM 325 [2,3] . P&CAM 234 [4,5] , a general colorimetric method for diazonium salts and azo dyes, has not been revised.
Source: Wikisource