National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Calibrate each personal sampling pump with a representative sampler in line. Sample at an accurately known flow rate between 1 and 4 L/min for a total sample size of 30 to 180 L. NOTE: This method will collect hydroquinone aerosol only. The equilibrium vapor pressure is equivalent to 0.11 mg/m 3 at 25 °C and may be a significant factor at elevated temperatures. Immediately after sampling, transfer the filter (do not include backup pad) with tweezers to a 60-mL ointment jar. Add 10 mL 1% acetic acid. Process field blanks similarly. Ship the filters in the ointment jars. ”