National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Calibrate each personal sampling pump with a representative sampler in line. Transfer 15 mL 0.1 M hydrochloric acid to a bubbler. Connect outlet arm of bubbler to the glass-wool-packed tube (to prevent splashover into the pump) and then to the sampling pump with the flexible tubing. Sample at an accurately known rate between 0.5 and 1.5 L/min for a total air sample of 3 to 20 L. Remove bubbler stem and rinse with 0.1 M hydrochloric acid into bubbler body. Seal bubbler with an inert stopper for shipment in a suitable container in order to prevent damage during transit. ”