“ Calibrate each personal sampling pump with a representative sampler in line. Transfer 10 mL 0.1 M hydrochloric acid to a bubbler. Connect outlet arm of bubbler to the glass-wool-packed tube (to prevent splashover into the pump) and then to the sampling pump with the flexible tubing. Sample at an accurately known rate of 0.2 to 1.0 L/min for total sample size of 2 to 100 L. Remove bubbler stem and rinse with 2 mL of 0.1 M hydrochloric acid into bubbler body. Seal bubbler with an inert stopper for shipment in a suitable container in order to prevent damage during transit. ”