National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Pyridine can cause liver and kidney damage, and CNS depression if it is inhaled or contacts the eyes or skin [4,5] . Methylene chloride is a suspect carcinogen [6] .
SAMPLING: 1. 2. 3. 4.
Calibrate each personal sampling pump with a representative sampler in line. Break the ends of the sampler immediately before sampling. Attach sampler to personal sampling pump with flexible tubing. Sample at an accurately known flow rate between 0.01 and 1 L/min for a total sample size of 18 to 150 L. Cap the samplers. Pack securely for shipment.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Sampler: glass tube, 7 cm long, 6-mm OD, 4-mm ID, flame-sealed ends, with plastic caps, containing two sections of activated (600 °C) coconut shell charcoal (front = 100 mg; back = 50 mg) separated by a 2-mm urethane foam plug. A silylated glass wool plug precedes the front section and a 3-mm urethane foam plug follows the back section. Pressure drop across the tube at 1 L/min airflow must be less than 3.4 kPa. Tubes are commercially available. Personal sampling pump, 0.01 to 1 L/min, with flexible connecting tubing.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Breakthrough (effluent concentration = 5% of test concentration) was not observed after sampling for 240 min at 0.93 L/min from an atmosphere containing 30.4 mg/m 3 pyridine in dry air. Carbon disulfide and methanol were tested and rejected as possible desorbing solvents. Carbon disulfide gave an average DE of 0.734; methanol, 0.201. Desorption efficiency using methylene chloride for 18 spiked samples in the range 0.8 to 3.1 mg pyridine per sample averaged 0.81 with S r = 0.013.
Source: Wikisource