National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Wear protective gloves and suitable protective clothing when handling the pure material [3] .
SAMPLING: 1. 2.
3. 4.
Calibrate each personal sampling pump with a representative sampler in line. Break the ends of the sampler immediately before sampling. Connect the two sorbent tubes with a short piece of flexible tubing. Attach sampler to personal sampling pump with flexible tubing. Sample at an accurately known flow rate between 0.01 and 0.2 L/min for a total sample size of 20 to 48 L. Separate the front and backup tubes. Cap the tubes with plastic (not rubber) caps.
”
Source: Wikisource

National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Sampler: two glass tubes, 10-cm long, 8-mm OD, 6-mm ID, flame-sealed ends, front tube containing a 100-mg section of extracted resin and a back containing a 50-mg section, retained by silylated glass wool plugs in each tube. Pressure drop across the tubes in series at 1 L/min airflow must be less than 3.4 kPa. (Tubes commercially available as SKC, Inc. ST 226-105) . 2. Personal sampling pump, 0.01 to 0.2 L/min, with flexible connecting tubing. 3. ”