National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ SAMPLE PREPARATION: 4. Open filter cassette; transfer filter to ointment jar. 5. Add 4.0 mL acetonitrile; gently swirl for 30 min. 6. Filter each sample using a 5-mL syringe with PTFE syringe filter or Swinney filtration device. Deliver filtrate to a 4-mL vial. CALIBRATION AND QUALITY CONTROL: 7. Prepare at least six working standards daily in the range 0.01 to 1 mg Rotenone per sample. a. Add known amounts of calibration stock solution to acetonitrile in 10-mL volumetric flasks and dilute to the mark. ”