“ Acetonitrile, HPLC grade.* Water, distilled, deionized. Thiram, reagent grade.* Air or nitrogen, compressed, for drying syringes. Calibration stock solution, 0.75 mg/mL. Dissolve an accurately weighed 7.5 mg Thiram in acetonitrile and dilute to 10 mL. Prepare fresh daily in duplicate. See SPECIAL PRECAUTIONS. 1. Sampler: 1-μm PTFE membrane filter (Millipore FA or equivalent) , 37-mm diameter, two-piece polystyrene cassette filter holder with backup pad, sealed with tape or a shrinkable band. Filter holders made of Tenite should not be used. 2. ”