National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Calibrate daily with at least six working standards over the range 2 to 20 μg antimony per sample. a. Add known amounts of antimony stock solution and conc. HCl to cover the antimony concentration of 2 to 20 μg Sb per 15 mL and a reagent blank. b. Proceed as in steps 10 through 15 and 18 and 19. c. Prepare a calibration graph (absorbance vs. μg antimony/15 mL) . Analyze three quality control spikes to ensure that the calibration graph is in control. ”