“ Add water to the 15-mL mark. CALIBRATION AND QUALITY CONTROL: 5. Calibrate daily with at least six working standards over the range 3 to 150 μg TMTU per sample. a. Pipet 0- to 10-mL aliquots of calibration stock solution into clean vials. Bring the total volume to 15 mL with distilled water. NIOSH Manual of Analytical Methods (NMAM) , Fourth Edition, 8/15/94 TETRAMETHYL THIOUREA: METHOD 3505, Issue 2, dated 15 August 1994 - Page 3 of 4 b. c. Analyze together with the samples and blanks (steps 6 through 8) . Prepare calibration graph (absorbance vs. μg TMTU) . ”