National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ The validated method was needed to provide air monitoring capabilities during methyltin chloride and stabilizer production. Table 2 lists the results of the validation experiments which followed the guidelines established by NIOSH [2] . However, fewer replicates and concentration levels were used than recomm ended. This method m et the NIOSH criteria for accuracy, bias and sample stability. The recom m ended sampling conditions are 250 ml/min for 5.5 hours maximum for time-weighted average (TW A) sampling and 1000 ml/min for 20 minutes maximum for short-term exposure limit (STEL) sampling. ”