Analyze the sample filters by step-scanning the most intense interference-free diffraction peak of zinc oxide and determine the integrated intensity. Measure the background on each side of the peak for one half the time used for peak scanning and add the counts from each side for a total (average) background. Determine the 2-theta position of the background for each sample. The net count or intensity, I x, is the difference between the peak integrated count and the total background count. The net intensity is normalized as in step 12 to obtain .
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Do not wash down the sides of the funnel after the deposit is in place to avoid disturbing it. Transfer the filter to the XRD sample mount. Perform step scans on the standards using the same diffraction peaks and instrumental conditions as for samples (step 16) . The resulting intensities for standards, , are normalized in the following procedure. Determine the net count, I r, of the reference specimen before or after each unknown, standard or blank is scanned. Use a diffraction peak of high intensity that can be measured rapidly but reproducibly (less than 1% S r) .