National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Determine the length of transfer line needed from the column end fitting to the bottom of the detector jet. Force a piece of 1.6-mm OD PTFE tubing equal to this length plus the length of the PTFE plug, through the center to the back end of the plug. Insert the PTFE plug, back end first, into the opening at the detector end of the column. Push the compression fitting over the column where the PTFE plug was inserted. Attach the column to the chromatograph, pushing the transfer line as close as possible to the bottom of the detector jet. Gently tighten the compression fitting around the column. ”