National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Determine desorption efficiency (DE) at least once per year for each lot of Tenax GC used for sampling in the range of interest. Determine for three samplers at each of five concentrations plus three media blanks. a. Transfer front sorbent section of media blank sampler to a 5-mL vial. b. Inject known amount (1 to 10 μL) of allyl glycidyl ether or standard solution of allyl glycidyl ether in hexane directly into sorbent in vial with a microliter syringe. NOTE: Standard solutions for DE determination should not contain internal standard. c. Cap the tube. Allow to stand overnight. ”