National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Remove and discard back sorbent section of a media blank sampler. Inject a known amount (1 to 20 μL) of calibration stock solution or a dilution thereof directly onto front sorbent section with a microliter syringe. c. Cap the tube. Allow to stand overnight. d. Desorb (steps 5 through 7) and analyze together with working standards (steps 11 and 12) . e. Prepare a graph of DE vs. mg methyl iodide recovered. Analyze three quality control blind spikes or three analyst spikes to ensure that the calibration graph and DE graph are in control. ”