National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Prepare calibration graph (phenyl glycidyl ether peak area vs. mg phenyl glycidyl ether per 0.5 mL) . Determine desorption efficiency (DE) at least once for each lot of charcoal used for sampling in the range of interest. Prepare three tubes at each of five levels plus three medial blanks. a. Remove and discard back sorbent section of a media blank sampler. b. Inject known amount (1 to 20 μL) of phenyl glycidyl ether or standard solution of phenyl glycidyl ether in CS 2 directly onto front sorbent section with a microliter syringe. c. Cap the tube. Allow to stand overnight. ”