“ Calibrate daily with at least six working standards. a. Add known amounts of calibration stock solution to CS 2 in 10-mL volumetric flasks and dilute to the mark. Prepare working standards in the range 10 to 100 μg vinylidene chloride/mL. b. Analyze with samples and blanks (steps 11 and 12) . c. Prepare calibration graph (peak area or peak height vs. μg vinylidene chloride) . Determine desorption efficiency (DE) at least once for each batch of charcoal used for sampling in the range of interest. ”