National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Rotator may also be substituted by a shaker or an agitator as long as the dissolution solution wets the wipe well. Sonication has also been shown to be effective. For dissolution of refractory materials such as high-fired beryllium oxide, agitation of the dissolution solution with the media must be replaced by heating to 80 °C for 30 minutes. Any standard dissolution process is particle-size dependent. ”