National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

The working range of this method is up to 100 mg/m3 for each element in a 500-L sample (the minimum range depends on the LOD for each sample; see Table 1) . The analysis is not compound specific. Certain elemental compounds are known to be acceptable or unacceptable by this method (see Table 3) . For unverified compounds, a test run should be conducted using a known amount of the compound in question to determine acceptability.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Value based on a 25-mL sample volume. The minimum sampling volume needed to obtain the OSHA PEL at the LOQ for the element/compound at a sample digestion volume of 25 m L. The m aximum sampling volume for a given sample, calculated by taking 50,000 :g as th e lim it for the element/compound per sample. NOTE: The LOD and LOQ values are dependent on the particular analytical instrument used.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

The method procedure is not suitable for samples at any weight with recoveries of between 90 and 110 percent. An alternative method should be used.
Valid up to 10,000 :g/sample and within 7 days of sample digestion. Va lid up to 50,000 :g/sample and at least 2 4 ho urs after s am ple digestion ; Valid up to 15,000 :g/s am ple within 24 hours of sample digestion. Valid up to 25,000 :g/sample and within 7 days of sample digestion. Valid up to 25,000 :g/sample and within 7 days of sample digestion. Valid up to 30,000 :g/sample and within 7 days of sample digestion.
Source: Wikisource