National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Prepare working standards by pipetting 0- to 10-mL aliquots of the 15 μg/mL standard solution into clean vials. Bring the total volume to 15 mL with distilled water. Follow steps 10 and 11 to analyze the working standards. c. Prepare a calibration graph (absorbance vs. μg per sample) . Determine recovery at least once for each lot of filters used. a. Place eight filters on a plastic test tube rack. Using an adjustable pipette, add to the center of each filter 0, 7.5, 15, 30, 45, 60, 90, 120, and 150 μL of 1000-μg/mL calibration stock solution. Let filters air dry overnight at room temperature. ”