National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Filter the sample solution through a syringe filter. CALIBRATION AND QUALITY CONTROL: 6. 7. Calibrate daily with at least six working standards over the range 1 to 70 μg strychnine per sample (0.2 to 14 μg/mL) . a. Add known amounts of calibration stock solution to mobile phase in 10-mL volumetric flasks and dilute to the mark. b. Analyze together with the samples and blanks (steps 9, 10, and 11) . c. Prepare calibration graph (peak area vs. μg strychnine) . Determine recovery (R) at least once for each lot of filters used for sampling in the calibration range. ”