National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ As an alternative to the glass fiber filters, air samples ofp-toluenesulfonic acid canbe collected in midget impingers containing isopropanol at 1 L/min. After sampling, the impinger solution is transferred to a 20-mL glass vial and transported to the laboratory. The vial is placed onto a heating plate maintained at 80C, and the isopropanol is evaporated to dryness with a gentle stream of nitrogen (total evaporation time is about 1.5 hours) . A 2-mL aliquot of 2% isopropanol in water is added to the vial, and the vial is placed into an ultrasonic bath for 30 seconds. ”