National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Calibrate daily with at least six working standards over the range 3 to 1200 μg F per sample. a. Add known amounts of calibration stock solution to 25 mL TISAB and 5 mL 20% NaOH (particulate fluoride) or 25 mL TISAB (gaseous fluoride) in 100-mL beakers. b. Adjust pH to 5.5 ± 0.2 with 1:1 HCl. c. Transfer to 50-mL volumetric flasks. Dilute to the mark with deionized water. NOTE: Working standards are stable at least one month in polyethylene bottles at 4 °C. d.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Calibrate each personal sampling pump with a representative sampler in line. Sample at an accurately known flow rate between 1 and 2 L/min for a total sample size of 12 to 800 L. Do not exceed 2 mg total dust loading on the filter. Pack securely for shipment.
SAMPLE PREPARATION: 4.
Cellulose ester membrane filter (particulate fluoride) . NOTE: Water-soluble particulate fluorides may be extracted by the procedure for gaseous fluorides. Insoluble particles require fusion. a. Transfer the membrane filter to a crucible containing 5 mL 20% NaOH. b. Evaporate to dryness. c.
Source: Wikisource