“ Prepare calibration graph (peak area or height vs. ng hexachloro-1,3-cyclopentadiene) . 9. Determine desorption efficiency (DE) at least once for each lot of Porapak T used for sampling in the range 25 to 12,500 ng hexachloro-1,3-cyclopentadiene per sample. Prepare three tubes at each of five levels plus three media blanks. a. Prepare five DE standard solutions in the range 5 to 2500 μg/mL by diluting aliquots of 20 mg/mL stock solution with hexane. b. Inject 5 μL of a DE standard solution directly into a media blank front sorbent section with a microliter syringe. ”