National Institute for Occupational Safety and Health

Summary

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Break the ends of the sampler immediately before sampling. Attach sampler to personal sampling pump with flexible tubing. Sample at an accurate flow rate between 0.01 and 0.2 L/min for a total sample size of 0.2 to 10 L. Cap the samplers with plastic caps and pack securely for shipment with bagged refrigerant.
SAMPLE PREPARATION: 5. 6. 7.
Place the front and back sorbent sections of the sampler tube in separate vials. Add 1.0 mL to CS 2 to each vial and attach a crimp cap to each vial. Allow to stand 30 minutes with occasional agitation.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Calibrate daily with at least six working standards over the range 2 to 2000 μg analyte per sample. a. Add known amounts of analyte to CS 2 in 10-mL volumetric flasks and dilute to the mark. b. Analyze together with samples and blanks (steps 11 and 12) . c. Prepare calibration graph (peak area vs. μg methyl acetate) . Determine desorption efficiency (DE) at least once for each lot of charcoal used for sampling in the calibration range (step 8) . Prepare three tubes at each of five concentrations plus three media blanks. a.
Source: Wikisource