National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Calculate concentration, C, analyte in the air volume sampled, V (L) : EVALUATION OF METHOD: Method S38 for methyl acrylate was issued on 12/6/74, and evaluated over a range of 13.9 to 58.4 mg/m 3 at an atmospheric temperature of 23°C and a pressure of 764 mm Hg, using 6-L samples [1] . Breakthrough was determined by sampling a generated atmosphere at a concentration of 59 mg/m 3 for 240 min at a flow rate of 0.190 L/min. Desorption efficiency (DE) was determined by spiking charcoal tubes at three levels, 0.087, 0.175, and 0.350 mg of methyl acrylate. ”