National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Add known amounts of calibration stock solution to carbon disulfide solvent in 10-m L volumetric flasks and dilute to the mark. b. Analyze together with samples and blank s (steps 11 and 12) . c. Prepare calibration graph (peak area vs. μg methyl acrylate) . 10. Determine the desorption efficiency (DE) at least once for each lot of Anasorb CMS used for sampling in the calibration range (ste p 9) . a. Prepare three tubes at each of five levels plus three media blanks. b. Inject a known amount of calibration stock solution directly onto the front sorbent section of each Anasorb CMS tube. ”