“ Spike a known amount of calibration stock solution directly onto the filter with a microliter pipet or syringe. b. Allow to stand overnight in the dark. c. Extract (steps 4 through 6) and analyze together with working standards (steps 10 and 11) . d. Prepare graph of recovery vs. mg o-terphenyl recovered. Analyze three quality control blind spikes and three analyst spikes to ensure that the calibration graph and recovery graph are in control. ”