National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Use of an internal standard is recommended to minimize errors caused by sample solvent evaporation and FPD response variations. c. Prepare a calibration graph of area vs. μg of tributyl phosphate per 10 mL of sample. Determine recovery in the concentration range of interest for each lot of filters used for sampling. Prepare three filters at each of five levels plus three media blanks. a. Spike aliquot of calibration solution onto each filter. b. After air-drying, extract filters with 10 mL diethyl ether (steps 5 through 7) . c. Analyze together with working standards (steps 11 and 12) . ”