National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Calibrate daily with at least six working standards over the range 1 to 300 μg CN - per sample. a. Prepare a working standard solution, 1.00 μg /mL.CN -, by diluting 100 μL of calibration stock solution to 100 mL with 0.1 N NaOH. b. Pipet 0.5-, 1.00-, 1.50-, 2.00- and 2.50-mL of the working standard solution into 25-mL volumetric flasks to make 0.50-, 1.00-, 1.50-, 2.00- and 2.50- μg CN - standards. c. Analyze together with field samples and blanks (steps 12 through 19) . d. Prepare calibration graph (absorbance vs. μg CN -) . ”
