Sodium trichloroacetate was used instead of trichloroacetic acid for all phases of this method development as well as in the preparation of standards. Trichloroacetic acid is very hygroscopic; the salt is much less so. 12.
Prepare working (calibration) standards by serial dilution to cover the analytical range. A suggested working standard concentration range is 0.08 to 300 μg/mL. Withdraw 200 μL of each calibration standard and follow steps 4 through 10. 13.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

There are numerous literature methods for the determination of TCAA in urine [13] . The National Center for Environmental Health/Centers for Disease Control and Prevention (NCEH/CDC) has a method that is more expensive but also significantly more sensitive [11] . This method is based on the procedure used by O’Donnell [7] with some modifications. NIOSH Manual of Analytical Methods (NMAM) , Fifth Edition TRICHLOROACETIC ACID IN URINE: METHOD 8322, Issue 1, dated 17 April 2015 - Page 2 of 5
REAGENTS:
EQUIPMENT:
1.
Sodium trichloroacetate [CAS #650-51-1] , 97% purity or greater* 2.
Source: Wikisource

National Institute for Occupational Safety and Health NIOSH Manual of Analytical Methods… (1994)

Calibrate daily with at least six liquid working standards covering the expected concentration range of the samples. MEASUREMENT: PRECAUTION: SYRINGE-RINSE SOLUTIONS: Toluene will extract material from some urine specimens that may eventually clog the syringe and cause injection errors unless the syringe is rinsed with the following solutions following each injection.