National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Dilutions can be stored in the freezer until determined that they are needed. CALIBRATION AND QUALITY CONTROL: 8.Calibrate daily with at least six working standards over the range 0.14 to 272 μg methyl bromide per sample from calibration stock solution. NOTE:0.085 μmoles bromine per μL is equivalent to a methyl bromide concentration of 8.07 μg/μL. a. Add known aliquots (2.0-, 20-, 45-, and 120-μL) of calibration stock solution to methylene chloride in 10-mL vials with PTFE-lined caps. Take 1 mL of lowest standard and dilute to 10 mL with methylene chloride. ”