National Institute for Occupational Safety and Health, NIOSH Manual of Analytical Methods… (1994)
“ Calibrate daily with at least six working standards. a. Add 10 mL of acidic permanganate solution and 1 mL of ferrous reagent to a 125-mL separatory funnel. b. Add 2 to 400 μL of the standard phosphate solution to cover the range 0.1 to 10 μg of PH 3. Add 8 to 9 mL of water to make the total volume of the solution (permanganate solution, ferrous solution, phosphate solution and water) equal to 20 mL. Prepare at least six calibration standards and a blank containing no phosphate. c. Add 7.5 mL of molybdate reagent and 25 mL of toluene-isobutanol solvent to the funnel. ”