“ Evaporate the acetylated sample to dryness under a gentle stream of nitrogen while warming to 60 °C on a hotplate. Redissolve the residue in 5.0 mL methanol, while agitating the sample in an ultrasonic water for 15 min. CALIBRATION AND QUALITY CONTROL: 10. 11. Prepare at least six working standards containing 0.01 to 4.0 μg/mL of the appropriate urea (s) (TDIU, HDIU, and/or MDIU) and 100 μg/mL of 1- (2-methoxyphenyl) piperazine by adding aliquots of calibration stock solutions to 2 mL methanol in a 10-mL volumetric flask. Add 10 μL acetic anhydride to each standard. ”