“ If possible, generate the correction curve data on the same day that coal mine dust samples are analyzed since curve param eters may vary somewhat from day to day. A curve through the points should be linear but will not pass through the origin since the peak at 915 cm ! 1 is more intense than the peak at 800 cm ! 1 and sm all am oun ts of kaolinite do not interfere with the quartz analysis. NOTE 1: Ka olinite correctio n calibration curve is required when s am ples are ashed in an LTA since kaolinite is not destroyed under these circum stances a nd has a s interfering peak at 800 cm ! ”
Source: Wikisource

National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Insert the sample into the beam of the spectrophotometer. Place the other half of the DM-450 filter on another holder, secure with a magnet and insert it into the reference beam. NOTE: For the best precision, the referenc e filter should be half of the same DM -450 filter which contains the redeposit. However, for routine analysis, the same 2-propanol-treated blank can be used for all filters with the sam e lot numb er. Run an infrared scan, in linear absorbance mode from 1000 to 650 cm ! 1. Draw an app ropriate baseline under the absorbance band at 800 cm ! 1 from ca. 820 to 670 cm ! 1. ”
Source: Wikisource

National Institute for Occupational Safety and Health,  NIOSH Manual of Analytical Methods… (1994)

“ Accuracy depends on obtaining uniform deposition of samples and standards across the filter surface and obtaining re prod ucible aliquo ts from the quartz s usp ens ion. This requires som e skill. The quartz calibration curve should be prepared before analyzing samples as a check on the analyst's ability to prepare uniform deposits. Repeatability should be <10% on re plicate standard s with m ore than 4 0 μg qua rtz. Perform an IR scan for each standard filter using the other half of the filter in the reference beam (steps 10 through 12) . ”
Source: Wikisource

Get perspective with Kwize: daily news enlightened by great literature